3.3 Open Collector Output
The Open collector output can be used as a switch that responds
when the flow rate moves above or below a setpoint, or it can be
used to generate a pulse that is relative to the flow volume or to
the flow rate.
Low
Output triggers when the flow rate is less than the setpoint.
The output will relax when the flow rate moves above the setpoint
plus the hysteresis value.
High
Output triggers when the flow rate is greater than the setpoint.
The output will relax when the flow rate drops below the setpoint
plus the hysteresis value.
Frequency
Output is a pulse stream that is based on the input flow sensor
signal. Set for 1 (input frequency = output frequency). Set for
even numbers (2, 4, 6, 8, . . . . 254 maximum) to scale output
frequency.
Pulse
Output is a pulse based on the volume of fluid that passes the
sensor. Set any value from 0.0001 to 99999.
The output may be disabled (Off) if not used.

VIEW menu

During normal operation, ProcessPro displays the VIEW menu.
When editing the CALIBRATE or OPTIONS menus, ProcessPro will return to the VIEW menu if no activity
occurs for 10 minutes.
To select a VIEW display, press the UP or DOWN arrow keys. The selections will scroll in a continuous loop.
Changing the VIEW display does not interrupt system operations.
No key code is necessary to change display selection.
Output settings cannot be edited from the VIEW menu.
View Menu
Flow1: 123.4 GPM
Flow2: 567.8 GPM
All of the displays below are temporary. After ten minutes the display will return to the permanent display.
Tot1: 1234567.8
Tot2: 123456.78>
Perm1: 1234567.8
Perm2: 123456.78
Loop 1 Output:
Loop 2 Output:
page 4 of 8
Display
Monitor the flow rate of Channel 1 and Channel 2 simultaneously.
This is the permanent view display.
Delta Flow:
Monitor the delta flow rate (Channel 1 rate - channel 2 rate = Delta Flow)
10.5 GPM
This is a permanent display.
Monitor channel 1 and channel 2 Resettable Totalizers. Press the RIGHT
ARROW key to reset the totalizer. If Reset is locked, you must enter Key
Code. Lock or Unlock function is in OPTIONS menu.
Monitor channel 1 Permanent Totalizer value.
Gallons
Monitor channel 2 Permanent Totalizer value.
Gallons
Monitor the 4-20 mA Loop 1 output.
12.00 mA
Monitor the 4-20 mA Loop 2 output.
12.00 mA
Last CAL:
Monitor date for scheduled maintenance or date of last calibration.
06-30-01
(See description in Calibrate Menu.)
Process
Hysteresis
Low Setpoint
Process
High Setpoint
Hysteresis
Description
‡ SIGNET
Relay energized
Relay relaxed
Flow1:
Flow2:
8550-3 Flow Transmitter
Time
Time
Flow
6.25 GPM
9.75 GPM
ENTER