Signet 8550-1 Flow Transmitter
*3-8550.090-1*
3-8550.090-1 Rev L 6/06 English
WARNING!
Remove power to unit before wiring input and out-
put connections.
Follow instructions carefully to avoid personal
injury.
1. Specifi cations
Dimensions
96 mm
(3.8 in.)
96 mm
(3.8 in.)
FRONT VIEW
Field Mount &
Panel Mount
General
Compatibility: Signet Flow Sensors (w/freq out)
Enclosure:
Rating:
NEMA 4X/IP65 front
Case:
PBT
Panel case gasket:
Neoprene
Window:
Polyurethane coated polycarbonate
Keypad:
Sealed 4-key silicone rubber
Weight:
Approx. 325g (12 oz.)
Display:
Alphanumeric 2 x 16 LCD
Update rate:
Contrast:
Display accuracy:
Thermal sensitivity shift: ±0.005% of reading per ºC
Electrical
Power:
Sensor Input:
Range: 0.5 to 1500 Hz
Sensor power:
2-wire: 1.5 mA @ 5 VDC ± 1%
3 or 4 wire: 20 mA @ 5 VDC ± 1%
Optically isolated from current loop
Short circuit protected
2. Installation
ProcessPro transmitters are available in two styles: panel mount and fi eld mount. The panel mount is supplied with the necessary
hardware to install the transmitter. This manual includes complete panel mounting instructions.
Field mounting requires one of two separate mounting kits. The 3-8051 integral kit joins the sensor and instrument together into a single
package. The 3-8050 Universal kit enables the transmitter to be installed virtually anywhere.
Detailed instructions for integral mounting or other fi eld installation options are included with the 3-8051 Integral kit or the 3-8050
Universal kit.
CALL 1-800-577-8111 FOR SALES AND SUPPORT
CLICK HERE TO RETURN TO WEBSITE
Optional
92 mm
Rear
(3.6 in.)
Cover
56 mm
41 mm
(2.2 in.)
)
(1.6 in.
97 mm
(3.8 in.)
SIDE VIEW
Panel Mount
1 second
User selected, 5 levels
±0.5% of reading @ 25ºC
12 to 24 VDC ±10%, regulated
61 mA max current
CALL 1-800-577-8111 FOR SALES AND SUPPORT
CLICK HERE TO RETURN TO WEBSITE
Contents
1. Specifi cations
2. Installation
3. Electrical Connections
4. Menu Functions
82 mm
(3.23 in.)
106 mm
(4.18 in.)
SIDE VIEW
Field Mount w/
8050 Universal base
Current output:
4 to 20 mA, isolated, fully adjustable and reversible
Max loop impedance:
Update rate:
Accuracy:
Open-collector output, optically isolated:
50 mA max. sink, 30 VDC maximum pull-up voltage.
Programmable for:
High or Low setpoint with adjustable hysteresis
Pulse operation (max rate: 300 pulses/min).
Environmental
Operating temperature: -10 to 70°C (14 to 158°F)
Storage temperature:
-15 to 80°C (5 to 176°F)
Relative humidity:
0 to 95%, non-condensing
Maximum altitude:
2000 m (6562 ft)
Insulation category:
II
Pollution degree:
2
Standards and Approvals
CE, UL listed
Immunity & Emissions:
Manufactured under ISO 9001 and ISO 14001
English
Flow 6.25 GPM
Total 1234567.8>
Signet Flow
Transmitter
96 mm
(3.8 in.)
102 mm
(4.0 in.)
SIDE VIEW
Field Mount w/
8051 Integral kit
50 Ω max. @ 12 V
325 Ω max. @ 18 V
600 Ω max. @ 24 V
100 ms
±0.03 mA
EN61326
ENTER

Advertisement

Table of Contents
loading

Summary of Contents for Signet 8550-1

  • Page 1 Field Mount w/ Panel Mount 8050 Universal base 8051 Integral kit General Current output: Compatibility: Signet Flow Sensors (w/freq out) • 4 to 20 mA, isolated, fully adjustable and reversible 50 Ω max. @ 12 V • Max loop impedance: Enclosure: 325 Ω...
  • Page 2: Electrical Connections

    Power + (BLACK) Terminals 1 and 2: AUXILIARY power Used only if the fl ow sensor requires more Internal open-collector than 1.5 mA current. For Signet sensors output circuit this is limited to the following products: Output -- • 2000 •...
  • Page 3 Auxiliary Power note: AUXILIARY power is used only if the flow sensor requires more than 1.5 mA current. For Signet sensors this is limited to the following products: 2000, 2507, 2551, open collector input signals. 3.2 Sensor Input Connections Transmitter Wiring Tips: Flow 6.25 GPM...
  • Page 4: View Menu

    Monitor the 4-20 mA Loop output. 12.00 mA Last CAL: Monitor date for scheduled maintenance or date of last calibration. (See description in Calibrate Menu.) 04-20-06 CALL 1-800-577-8111 FOR SALES AND SUPPORT Signet 8550-1 Flow Transmitter CLICK HERE TO RETURN TO WEBSITE...
  • Page 5 fl ashing. This will recall the Output Setpnt: Saving last saved value of the item being edited and return you to Step 3. Output Setpnt: 19.00 GPM > CALL 1-800-577-8111 FOR SALES AND SUPPORT Signet 8550-1 Flow Transmitter CLICK HERE TO RETURN TO WEBSITE...
  • Page 6: Calibrate Menu

    (2, 4, 6, 8 . . . 254 maximum) to scale output frequency. Divide by 1 > Use this “note pad” to record important dates, such as annual recertifi cation or scheduled Last CAL: maintenance. 4-20-06 CALL 1-800-577-8111 FOR SALES AND SUPPORT Signet 8550-1 Flow Transmitter CLICK HERE TO RETURN TO WEBSITE...
  • Page 7: Options Menu

    60 s 70 s very quickly. (Solid blue line) NOTE: The SENSITIVITY function is ineffective if the AVERAGING function is set to zero (seconds). CALL 1-800-577-8111 FOR SALES AND SUPPORT Signet 8550-1 Flow Transmitter CLICK HERE TO RETURN TO WEBSITE...
  • Page 8: Troubleshooting

    Heavy duty wall mount bracket 3-5000.598 198 840 225 Surface Mount Bracket 3-5000.399 198 840 224 5 x 5 inch adapter plate for Signet retrofi t 3-9000.392 159 000 368 Liquid tight connector kit for rear cover (includes 3 connectors) 3-9000.392-1 159 000 839 Liquid tight connector kit, NPT (1 piece) 3-9000.392-2...