3.3 Relay Output
The relay output can be used as a switch that responds when the
process value moves above or below a setpoint.
Low:
Output energizes when process variable is less than the setpoint.
The output will be de-energized when the process variable moves
above the setpoint plus the hysteresis value.
High:
Output energizes when process variable is greater than the
setpoint. The output will be de-energized when the process
variable moves below the setpoint plus the hysteresis value.
Off
The relays may be disabled if not being used.

VIEW menu

During normal operation, the ProcessPro displays the VIEW menu.
When using the CALIBRATE or OPTIONS menus, the ProcessPro will return to the VIEW menu if no
activity occurs for 10 minutes.
To select the item you want displayed, press the UP or DOWN arrow keys. The items will scroll in a
continuous loop.
Changing the display selection does not interrupt system operations.
No keycode is necessary to change display selection.
Output settings cannot be edited from the VIEW menu.
View Menu
The displays below are temporary. After 10 minutes the display returns to the permanent display.
4
Display
Monitor the pressure input from the sensor.
Pressure:
60 PSI
This is the permanent display.
Loop Output:
Monitor the 4 to 20 mA Loop output.
12.00 mA
Last CAL:
Monitor date for scheduled maintenance or date of last calibration.
02-10-09
(See description in Calibrate Menu.)
Process
Hysteresis
Low Setpoint
Output energized
Output de-energized
Process
High Setpoint
Hysteresis
Description
Time
Time
Pressure Cal:
0.0
psi >
Signet Pressure
Transmitter
Relay 1
Relay 2
ENTER
Signet 8450-2 Pressure Transmitter