NTRODUCTION AND
SPECIFICATIONS
1-16.
MNEMONICS
The mnemonics
used
in
the schematics,
block diagrams, wiring diagrams,
truth
tables,
and
the
text
are
listed in
Figure
8-1.
1-17.
SIGNAL
GENERATOR
SPECIFICATIONS
Unless otherwise noted,
the
following
performance
is
guaranteed over
the specified
environmental
and
ac
power
line
conditions 20
minutes
after
turn-on.
Table
1-1
lists
the
Generator
specifications.
Table
1-1.
6062A
Synthesized
RF
Signal
Generator
Specifications
Warranted performance^
20
minutes after power-on within operating temperature
range.
FREQUENCY
(9
1/2-Digit Display)
RANGE
0.1
to
2100.0
MHz
in 4
bands;
0.1
to
244.99999
MHz,
245 to
511.99999
MHz,
512 to
1049.99999
MHz,
1050 to
2100.0
MHz.
RESOLUTION
... 10
Hz
from 0.1 to 1050 MHz,
20
Hz from 1050 to 2100
Itiz.
ACCURACY
Same as
reference
(See
REFERENCE).
REFERENCE (Internal)
The unit
operates
on an internal
free-air 10-MHz crystal oscillator
with
an
ageing
rate
of
<t0.5
ppm/month and
<+5 ppm for 25°C,
+25°C, Internal reference signal
(10-Miz)
available
at rear
connector,
level >0
dBm, terminated
in
50-ohms.
REFERENCE (External)
Accepts
1-, 2-, 2.5-, 5-, or
10-MHz
signal. Level
of
0.3 to
4.0 Vpp
into
50-ohms termination.
AMPLITUDE
(3
1/2-Digit Display)
RANGE (Indicated)
+16 (+16
peak with
AM
enabled) to -137
dBm,
from 0.1
to 1049.99999
MHz.
+13 (+13
peak with
AM enabled) to
-137
dBm,
from
1050
to
2100
MHz.
(Autoranging 6-dB
step
attenuator).
RESOLUTION
0.1
dB (<1% or
1
nV
in
volts).
Annunciators
for
dB,
dBm,
V,
mV, uV,
dB mV,
and
dB
uV.
ACCURACY
+1
dB from
+16
to
-127 dBm from
?
to
1049,99999
MHz.
*
+1.5 dB from
+13
to
-127 dBm from
?050 to 2100
mz.
+2.0
dB from
+16
to
-127 dBm from
0.1
to
.99999
MHz.
*
+1.5 dB
at
temperatures other
than
23°C +5°C
1-5