Rf generators

Advertisement

Tabie
2-10.
IEEE-488
Address
Mode
Message
Descriptions
(cont)
1
MESSAGE
DESCRIPTION
Listener
The
Generator unlistens
and untalks when the
IFC
message
Operation
is
received.
OTA
Other
Talk
Address
Talker
Operation
The
Generator untalks when
the
OTA message
is
received.
SPE
Serial Poll
Enable
Talker
After receiving
the SPE
message, the Generator responds
Operation
with the serial poll status byte,
if
addressed
to talk.
SPD
Serial Poll Disable
Talker
After receiving
the SPD
message, the Generator resumes
Operation
normal talk operation.
ULA
Unlisten Address
Listener
The
Generator unlistens when
the
ULA message
is
Operation
received.
RSV
Request Service
Service
The front panel SRQ indicator
is
Lit
when
the rsv
Operation
message
is
sent. The Generator may request service
for
several reasons. Each reason
for
service request
can be
individually masked with
the set
mask command
("SM"). The service request mask can be
interrogated with
the
interrogate mask command
("IM").
DCL
Device Clear
Clear
The DCL
message
is
ignored when
in
local.
When
the DCL
Operation
message
is
received (during
remote
operation)
the
Generator
is
cleared. Any characters
in
the input buffer
are
cleared followed
by the same
operation
as the clear
command ("CL"). The operation
of
the DCL
message
is
identical
to the
operation
of
the
selected device clear
(SDC)
message.
The
cleared state
of
the
Generator
is
described
in
the
paragraphs
on
Power-On Conditions.
SDC
Selected Device
Clear
Clear
The SDC
message
is
ignored during Local operation.
Operation
When the SDC
message
is
received (during
rmote
operation), the Generator
is
cleared. Any
characters
in
the input
buffer are cleared followed
by the
same
operation as
the clear command ("CL"). The
operation
of the
SDC
message
is
identical to the operation
of
the
device
clear (DCL) message. The cleared state
of
the
Generator
is
described
in
the paragraphs on
Power-On Conditions.