Fluke 6062A Instruction Manual page 49

Rf generators

Advertisement

Dev
i
ce
CL
none
none
Clears the instrument state
Clear
and
exits
all
compensation
procedures.
COMPENSATION MODE COMMANDS
Compensation
CM
none
AT
Begin Attenuator compensation
Mode
FM
procedure.
Begin
FM
compensation procedure.
OD
Begin Output with default
attenuator compensation procedure.
OT
Begin Output
compensation
procedure.
sv
Save
compensation
data and exit
compensation procedure.
EX
Exit
compensation procedure
without
saving the data.
COMPENSATION PROCEDURE COMMANDS
Compensa
ti
on
CF
float
G2
Accept
FM
deviation reading
FM
Entry
MZ
from
modulation analyzer
KZ
during
FM
compensation
HZ
procedure.
Compensati on
CP
float
DB
Accept
amplitude reading from
Ampl
i
tude
power meter
during level
Entry
compensati on procedure.
Interrogate
IC
none
FR
Interrogate the frequency
of
Compensation
the
current compensation
Step
Data
TG
procedure
step.
Interrogate
the
target
level
or
FM
deviation
of the
current
compensation procedure
step.
Interrogate
IH
none
none
Interrogate
the
frequency
and
Het Adjust-
level
offset necessary
to make
ment Data
the het
compensation adjustment.
For
example,
"+00001
20000 .,+00000000 .30"<E0R>
indicates the adjustment should
be
made
at
120 kHz and the
Level
needs to
be
adjusted
up
.3 dB.
EDIT ENTRY
Position
AB
float
V
Position the bright digit
in
Amplitude
MV
the
AMPLITUDE display with
Bright
Digit
uv
the
stated resolution.
For
NV
example, enter "ABlOMV"
for
DB
a
10-mV resolution.
Position
DB
float
GZ
Position the bright digit
in
FM
Bright
MZ
the
FM
display
with the stated
Digit
KZ
resolution.
For
example, enter
HZ
"DB1KZ"
for
a
1-kHz resolution.
RD