To Change The Logic Threshold For Digital Channels - Agilent Technologies 54621A User Manual

Hide thumbs Also See for 54621A:
Table of Contents

Advertisement

To change the logic threshold for digital channels

1 Press the
D15 Thru D8
2 Press the
Threshold
to define your own threshold.
User
Logic family
Threshold Voltage
TTL
1.4 V
CMOS
2.5 V
ECL
–1.3 V
User
Variable from –8 V to +8 V
The threshold you set applies to all channels within the selected D15 Thru D8 or
D7 Thru D0 group. Each of the two channel groups can be set to a different
threshold, if desired.
Values greater than the set threshold are high (H) and values less than the set
threshold are low (L).
3 When the
Threshold
the Entry knob to set the logic threshold voltage for the channel group.
MegaZoom Concepts and Oscilloscope Operation
To change the logic threshold for digital channels
key or
D7 Thru D0
key.
softkey, then select a logic family preset or select
softkey is set to
User
, press the
User
softkey, then turn
4-37

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

54622a54642a54621d54622d54624a54641d ... Show all