To set up the I/O port to use a controller

Use the I/O menu to make settings if you have a controller connected to the
oscilloscope.
Controller
RS-232
type
Baud rate
1 Press the
Controller
GPIB.
GPIB will be a selection on this softkey only if you have the optional N2757A
GPIB Interface Module connected to the oscilloscope.
The RS-232 configuration to the connector on the rear panel is:
Data bits - 8
Stop bits - 1
Parity - none
2 Press the
XON DTR
The RS-232 handshake protocol can be set to:
XON XON (transmit on)/XOFF (transmit off), a software handshake, or
DTR data terminal ready, a hardware handshake.
The default handshake is DTR.
3 Press the
softkey to set the RS-232 baud rate.
Baud
The RS-232 baud rate can be set to 9600, 19200, 38400, or 57600.
The default baud rate is 9600.
4 If you have the optional N2757A GPIB Interface Module connected to
the oscilloscope, turn the Entry knob to set the GPIB address shown in
the
Address
softkey.
The Address softkey is only displayed if the GPIB module is attached to the
oscilloscope.
Turn the Entry knob to set the GPIB address of this oscilloscope. The
oscilloscope address can be set to any address from 0 to 30. The default
oscilloscope address is 7.
To set up the I/O port to use a controller
Handshake
GPIB
address
softkey to set the controller interface to RS-232 or
softkey to set the RS-232 handshake.
Utilities
Return to
previous menu
6-9