Table 3-10 Dcv And Acv Auto Resolution - Keithley 2001 Operator's Manual

Hide thumbs Also See for 2001:
Table of Contents

Advertisement

Front Panel Operation
Table 3-9
DCV and ACV auto filter
Measurement function and
type
DCV
DCV peak spikes
RMS, average, low frequency
RMS
ACV peak
RESOLUTION
Except for frequency, temperature, and some special cases of
AC voltage, all functions can operate with 3.5, 4.5, 5.5, 6.5,
or 7.5-digit resolution, or they can default to a setting appro-
priate for the selected integration time. You can program the
resolution parameter as follows:
1. From the normal reading display, press the CONFIG
key and then the appropriate function key to access the
top level of a function configuration menu. For example,
the CONFIGURE DCV menu is displayed as follows:
CONFIGURE DCV
SPEED FILTER RESOLUTION
Table 3-10
DCV and ACV auto resolution
Measurement function and type Integration time
DCV
DCV peak spikes
RMS, average
Low frequency RMS
ACV peak
Notes:
1. If the DCV resolution is AUTO and the integration time SET-BY-RSLN, the resolution will be 6.5 digits and the
integration time 1.0 PLC.
2. For DCV peak spikes, low frequency RMS, and ACV peak measurements, the integration time setting is ignored.
3. For RMS and average measurements, if the resolution is AUTO and the integration time is SET-BY-RSLN, the res-
olution will be 5.5 digits and the integration time 1.0 PLC.
3-18
Units
State
Type
-
On
Advanced
-
On
Advanced
Any
Off
Advanced
Volts
On
Advanced
dB, dBm
On
Averaging
0.01 to <0.02 PLC
0.02 to <0.20 PLC
0.20 to <2.00 PLC
2.00 to 10.00 PLC
Not used
0.01 to <0.02 PLC
0.02 to <10.00 PLC
10.00 PLC
Not used
Not used
Noise
Readings
tolerance
10
1.0%
10
5.0%
10
5.0%
10
5.0%
10
-
2. Use the cursor keys (
TION and press ENTER. The following typical menu is
shown:
SET DCV RESOLUTION
AUTO 3.5d 4.5d 5.5d 6.5d 7.5d
3. Highlight the desired resolution and press ENTER.
The AUTO selection optimizes the resolution for the present
integration time setting. See Table 3-10 for the default reso-
lutions of the DCV and ACV functions. The default resolu-
tions of other functions are listed in paragraphs 3.4.2, 3.4.3,
and 3.4.5.
Resolution
4.5d
5.5d
6.5d
7.5d
3.5d
4.5d
5.5d
6.5d
5.5d
4d
Averaging
Mode
Moving
Moving
Moving
Moving
Moving
and
) to highlight RESOLU-

Hide quick links:

Advertisement

Table of Contents
loading