Site preparation overview
Introduction
Before the product is delivered, you must prepare the site according to the instructions
in this chapter.
Select the site
Select a site that complies with national and international requirements for biosafety.
General site requirements
The following requirements must be fulfilled at the location where the ÄKTApilot system
will be installed:
Space requirements
Prepare a stable laboratory bench with a height of 1200 mm. The following space re-
quirements must be fulfilled at the location where the ÄKTApilot system will be installed:
Parameter
Minimum bench area for placement of
the ÄKTApilot system
Recommended minimum bench area for
operating the ÄKTApilot system
ÄKTApilot Site Preparation Guide 29-1396-54 AA
WARNING
Spread of biological agents. The operator must take all necessary
actions to avoid spreading hazardous biological agents. The facility
must comply with the national code of practice for biosafety.
ÄKTApilot is intended for indoor use only
The instrument units must not be installed in a dusty environment
The instrument units must not be exposed to direct sunlight
2 Site preparation overview
Specification
1200 x 600 mm
2000 x 800 mm
5