Invacare® Platinum™ Mobile
Surge
± 1 kV line(s) to line(s)
IEC 61000-4-5
± 2 kV line(s) to ground
0% U
Voltage dips
and
IEC 61000-4-11
70% U
Single phase: at 0°
Voltage Interruptions
0% U
IEC 61000-4-11
Power frequency
(50/60 Hz) magnetic
30 A/m
field
IEC 61000-4-8
66
± 1 kV line(s) to line(s)
± 2 kV line(s) to ground
; 1 cycle
0% U
T
and
; 25/30 cycles
70% U
T
Single phase: at 0°
; 250/300 cycle
0% U
T
30 A/m
Mains power quality should be that of a typical commercial
or hospital environment.
Mains power quality should be that of a typical commercial
or hospital environment. If the user of the Device requires
; 1 cycle
T
continued operation during power mains interruptions, it
is recommended that the Device be powered from an
; 25/30 cycles
un-interruptible power supply or a battery.
T
U
is the a. c. mains voltage prior to application of the
T
test level.
Mains power quality should be that of a typical commercial
or hospital environment. If the user of the Device requires
continued operation during power mains interruptions, it
is recommended that the Device be powered from an
; 250/300 cycle
T
un-interruptible power supply or a battery.
U
is the a. c. mains voltage prior to application of the
T
test level.
Power frequency magnetic fields should be at levels
characteristic of a typical location in a typical commercial
or hospital environment.
1194969-A-00