Invacare® Platinum™ Mobile
LCD DISPLAY TEXT:
Shut Down
Battery Cold
INDICATORS:
Two audible beeps every 16 seconds
YELLOW alarm indicator illuminated
NOTES:
To clear the alarm, turn the concentrator off. Complete solutions 1, 2, 3, or 4 before restarting.
LCD DISPLAY TEXT:
Shut Down
Oxygen Fault
INDICATORS:
Two audible beeps every 16 seconds
YELLOW alarm indicator illuminated
NOTES:
To clear the alarm, turn the concentrator off. Complete Solution 1 and/or 2 before restarting. The alarm limit has a tolerance of +/- 2%
Alarm signal generation may be delayed by up to 128 seconds due to the monitoring algorithm used to prevent nuisance alarms.
Call your provider and report the problem if alarm continues.
58
DESCRIPTION:
Battery is too cold to allow the battery to
continue operating.
Concentrator function has shut down.
DESCRIPTION:
The oxygen output purity has fallen to a value
below 73%.
Concentrator function has shut down.
SOLUTIONS:
1.
Move concentrator to warmer
surroundings. Allow concentrator to
warm up to 41°F (5°C), OR
2.
Use AC or DC power and remove
battery, OR
3.
Insert an alternate charged battery and
remove the cold battery, OR
4.
Change to another source of oxygen if no
other power source is available.
SOLUTIONS:
1.
Verify concentrator is in recommended
environmental temperature. Refer to
Specifications in Technical Data.
2.
Clean intake filter and ensure both intake
and exhaust are not blocked.
3.
Change to another source of oxygen if
alarm continues.
1194969-A-00