CAUTION!
Risk of Damage
– Use of sterilization or cleaning solutions may leave
this device nonfunctional.
CAUTION!
Risk of Minor Injury or Discomfort
To ensure proper breath detection and oxygen delivery:
– Ensure the cannula prongs are positioned properly in
your nose.
– DO NOT use tubing/cannula length exceeding 25 ft
(7,6 m).
– Use crush-proof oxygen tubing.
– Use ONLY single lumen cannulas with the flow
characteristics stated in Specifications in Technical
Data.
– Avoid use of concentrator in windy conditions.
– Check for gas flow at the outlet of the cannula.
1194969-A-00
1.
Remove the cannula from its packaging.
2.
Connect the cannula to the concentrator oxygen outlet port A.
Ensure the connection is secure.
Place the cannula B over your ears and position the prongs in
3.
your nose as instructed by your health care provider or cannula
manufacturer.
Gas flow at the outlet of the cannula can be checked while
the concentrator is warming up. Wave your hand in front
of the nasal prongs. You should be able to hear and feel the
pulsed flow of gas. If you do not feel the gas pulse, check the
cannula connection for leaks.
Setup
31