Invacare® Platinum™ Mobile
Concentrator is Beeping with No Lights or Display
Symptom
Concentrator is beeping with no
lights or display.
Beeping will last less than
two minutes if no power
source is provided.
Battery Not Charging When External Power is Connected
Symptom
With AC or DC
Power source is faulty, or there is a loose connection.
power plugged into
the concentrator,
Battery is outside the allowed temperature range for
the external power
charging.
icon is illuminated
and battery charge
Battery is not fully seated.
level indicator is not
Battery charge is complete (all four segments of charge
flashing.
level indicator are illuminated).
46
Probable Cause
Battery has been removed or disconnected and
there is no external power connected.
When operating without a battery, external power
is lost.
Probable Cause
Solution(s)
1.
Ensure the battery is completely inserted and
restart.
2.
Install a different charged battery and restart.
1.
Try another power outlet and check connections
to concentrator and restart.
Solution
1.
Try another power outlet and check connections to
concentrator.
1.
Allow concentrator and battery to cool down to less
than 104°F (40°C), or warm up to 41°F (5°C).
1.
Ensure the battery is completely inserted.
1.
No action required.
1194969-A-00