LCD DISPLAY TEXT:
Low Battery
Charge Concentrator
INDICATORS:
Single audible beep no repeat
YELLOW alarm indicator illuminated
Battery gauge:
Pulse Flow Settings P1–P3: only the
lowest segment is illuminated
Pulse Flow Setting P4: only the lowest
two segments are illuminated
NOTES:
The alarm turns off when external power is connected or when a charged battery is installed.
Alarm will turn off when a breath is detected.
LCD DISPLAY TEXT:
Low Battery
Charge Now
INDICATORS:
Single audible beep every 30 seconds
YELLOW alarm indicator illuminated
Battery gauge:
Only the lowest segment is illuminated
1194969-A-00
DESCRIPTION:
Remaining battery run time of approximately
30 minutes. Battery requires charging.
DESCRIPTION:
Remaining battery run time of approximately
15 minutes or less. Battery requires charging.
Loss of power is imminent.
SOLUTIONS:
1.
Connect oxygen concentrator to either
the AC or DC power or insert an
alternate charged battery into the unit,
OR
2.
Insert an alternate charged battery into
the unit. Remove the depleted battery
and charge it with the external battery
charger for future use.
SOLUTIONS:
1.
Connect oxygen concentrator to either
the AC or DC power, OR
2.
Insert an alternate charged battery into
the unit. Remove the depleted battery
and charge it with the external battery
charger for future use, OR
3.
If no other power source is available,
change to another source of oxygen.