Invacare® Platinum™ Mobile
LCD DISPLAY TEXT:
Low Oxygen
See Manual
INDICATORS:
Single audible beep every 30 seconds
YELLOW alarm indicator illuminated
NOTES:
The alarm will turn on when the oxygen purity has dropped below 85%.
The alarm will turn off when the oxygen output purity has increased above to 87%.
The alarm limits have a tolerance of +/- 2%.
Alarm signal generation may be delayed by up to 128 seconds due to the monitoring algorithm used to prevent nuisance alarms.
Call your provider and report the problem if the warning continues with next use.
LCD DISPLAY TEXT:
Cannot charge
Allow to cool
INDICATORS:
Single audible beep every 30 seconds
YELLOW alarm indicator illuminated
NOTES:
The alarm will turn off when the required battery temperature is reached, when the battery is removed, or when external power is disconnected.
52
DESCRIPTION:
The oxygen output purity has fallen to a value
between 73% and 87%.
DESCRIPTION:
Battery is too hot to allow the battery to
charge.
This alarm only occurs when External Power
is connected.
SOLUTIONS:
1.
Verify concentrator is in recommended
environmental temperature. Refer to
Specifications in Technical Data.
2.
Clean intake filter screen and ensure both
intake and exhaust are not blocked.
3.
Turn concentrator off, then on again to
retry.
SOLUTIONS:
1.
Move concentrator to cooler
surroundings. Allow concentrator
to cool down to less than 104°F (40°C),
OR
2.
Remove battery (with AC or DC power
connected) to allow battery to cool.
1194969-A-00