Invacare® Platinum™ Mobile
LCD DISPLAY TEXT:
Warming Up
INDICATORS:
No audible beep
YELLOW alarm indicator illuminated
NOTES:
Depending on a variety of factors, it can take up to 15 minutes to reach the minimum specified oxygen purity.
The alarm will turn off when the minimum oxygen output purity is reached.
If the minimum oxygen output purity has not been reached within 15 minutes from initial start up, the alarm will become a Low Purity alarm
condition.
54
DESCRIPTION:
The concentrator is still warming up after
two minutes from initial start up. The oxygen
output purity has not reached the minimum
value listed in Specifications in Technical Data.
SOLUTIONS:
1.
Verify concentrator is in recommended
environmental temperature. Refer to
Specifications in Technical Data.
2.
Clean intake filter screen and ensure both
intake and exhaust are not blocked.
3.
Allow concentrator to continue warming
up for at least 15 minutes.
4.
Turn concentrator off, then on again to
retry.