Disinfect The Internal Water Circuit And Hoses Every 14 Days; Draining The Internal Water Circuit And Hoses For Disinfection - Stryker altrix Operation Manual

Precision temperature management system
Hide thumbs Also See for altrix:
Table of Contents

Advertisement

Disinfecting external surfaces (Continued)
Graphical user interface display
Handle
14. Follow specified contact time in accordance with the disinfectants manufacturer's instructions for use.
15. To rinse, wipe the hoses and patient temperature probe cables, with a soft, lint free cloth moistened with sterile
distilled water.
16. To rinse, wipe the controller, reservoir and reservoir lid surfaces, and system components with a lint free cloth
moistened with sterile distilled water.
17. To dry, wipe the controller, reservoir, reservoir lid surfaces, and system components with a clean, dry cloth to
remove any excess liquid.
18. Replace the reservoir.
19. Allow the external surfaces of the controller and components to dry thoroughly.
20. Store the power cord, cables, and hoses.

Disinfect the internal water circuit and hoses every 14 days

Use the BruClean TbC disinfectant tablets by BruClean TbC (EPA registration number 71847-2-106) before first use, at
least every 14 days, and before storage. BruClean TbC has been validated for internal water circuit disinfection. Make
sure that you follow the disinfectant manufacturer's guidelines to avoid the risk of injury. Failure to follow the
disinfectant's instructions may void your warranty.
CAUTION
Always use sterile distilled water or water that has been passed through a filter less than or equal to 0.22 microns
with this product.
Do not disinfect the internal water system with a thermal transfer device attached as this may cause a leak.
Do not use bleach or any other cleaning or disinfectant agents for internal circuits. This could result in damage to the
product. Only use approved disinfectant tablets.
Always drain the product before disinfecting the internal water circuit. Failure to drain the product may reduce the
effectiveness of the disinfection process.
Note: Disinfection of the Altrix internal water system was validated using M. mucogenicum.
Tools Required:
2 gallons (7.6 L) of sterile distilled water or water that has been passed through a filter less than or equal to 0.22
microns
Personal protection equipment (PPE) as recommended by the disinfectant manufacturer's instructions
Soft, lint free cloth (2 or more)
2 BruClean TbC 13.1 g tablets (Active ingredient NaDCC solution ppm = 1874 mg/L)
Note: BruClean TbC is a blend of 48% sodium dichloroisocyanturate and Adipic Acid with a 5% sodium dodecyl
benzene sulphonate surfactant.
Service tool adapter hose (8001-999-017) for Colder style connector hoses
Floor drain
See
Product illustration on page 10

Draining the internal water circuit and hoses for disinfection

1.
Unplug the power cord from the wall outlet.
2.
Place the controller over a floor drain.
www.stryker.com
Disinfecting
for clarification of product component names and locations.
8001-009-001 REV G
39