Intended use
The cleaning programme must be chosen according to the type of
soiling and load being processed.
The agents used for reprocessing the products should be selected to
suit the level of cleaning required.
The cleaning result must ensure that items can be disinfected
correctly, that subsequent sterilisation can be carried out and that the
items can be used again safely.
Medical devices are best processed using the Vario TD programme
where applicable.
The use of a suitable carrier (mobile unit, module, insert, etc.) is
important to ensure the adequate cleaning of the load. Examples are
given in the section "Areas of application".
This machine is programmed to carry out the final rinse with mains
water or with processed water of a quality to suit the application (e.g.
purified water, fully demineralised water or demineralised water).
This machine complies with EN ISO 15883 for validation purposes.

Spray pressure and spray arm monitoring

This machine has a sensor for monitoring wash pressure during
active cleaning process phases.
Spray arm pressure monitoring is carried out in accordance with the
general validation guidelines of the German Society for Hospital
Hygiene (DGKH), the German Society for Sterile Supply (DGSV) and
the Working Group for Instrument Preparation (AKI) for the validation
and routine monitoring of machine cleaning and disinfection
processes for thermally stable medical devices in accordance with
EN ISO 15883.
Wash pressure monitoring results are recorded in the process
documentation.
Spray arm speed can also be monitored, e.g. for prompt detection of
blockages due to misloading or foam in the water circulation system.
Spray arm monitoring can be activated or deactivated by Miele
Service.