Process validation

The standard of cleaning and disinfection in the disinfection
programmes must be confirmed by the user as a routine matter.
Safety checks and performance validation must be carried out in
accordance with the internationally recognised norm EN ISO 15883.
In some countries national regulations guidelines and
recommendations also apply.
In the UK these are:
– The Medical Devices Directive (MDD)
– The Medical Devices Operation Ordinance
– National Health and Safety regulations regarding disinfection,
CFPP, HTM or relevant technical memorandums (depending on the
machine) in the UK.
The sensor test point for validation is located at the front right on the
top of the machine under the lid or the worktop. To reach the access
point, the lid of the machine must be removed or the machine must
be pulled out from under the worktop.
Open the door.
Unscrew the fixing screws.
Then remove the safety screws on the back of the machine from
the lid and lift the lid to remove it.
Or
Pull the washer-disinfector out by approx. 15 cm from under the
worktop.