Spray Pressure And Spray Arm Monitoring - Miele PG 8593 Operating Instructions Manual

Washer-disinfector for laboratory glassware and laboratory utensils
Table of Contents

Advertisement

Operation

Spray pressure and spray arm monitoring

The machine has a sensor for monitoring spray pressure in order, for
example, to detect pressure fluctuations due to misloading or foam in
the water circulation system. Spray pressure monitoring is set at the
factory to active in the "Cleaning" and "Final rinse" wash blocks.
The spray pressure monitoring result is documented together with
process documentation.
Spray arm speed can also be monitored, e.g. for detection of
blockages due to misloading or foam in the water circulation system.
Spray arm speed monitoring can be activated or deactivated via the
programmable settings.
Miele Service can make further settings for spray pressure and spray
arm monitoring.