Operating Theatre Shoes - Miele PG8582 CD Operating Instructions Manual

Washer-disinfector
Hide thumbs Also See for PG8582 CD:
Table of Contents

Advertisement

Operating theatre shoes

 Theatre shoes should be cleaned and disinfected in a machine
installed specifically for this purpose only. This is to ensure, for
example, that any fluff or soiling cannot settle inside the lumen of
hollow instruments.
Theatre shoes can only be reprocessed together with other items if
a risk assessment has been carried out by the user.
Heat-sensitive theatre shoes and insoles should be cleaned and
chemo-thermally disinfected at 60 °C. A special programme has to be
uploaded by Miele Service, and an additional dispensing system for
dispensing chemical disinfectant must be retrofitted.
To establish the effectiveness of chemical disinfection agents for
disinfection please consult the manufacturer of the chemical agent.
A thermal disinfection programme
as the manufacturer has confirmed that the theatre shoes are thermo-
stable to 80 °C.
Remove insoles before reprocessing theatre shoes.
Please equip the upper and lower carriers with the following inserts
for reprocessing theatre shoes:
– A 151 with A 307 insert for theatre shoes up to size 48 (Eur).
– A 101 with A 308 insert for insoles up to size 45 (Eur).
A large amount of fluff can build up in the machine when cleaning
theatre shoes. The filters in the wash cabinet should therefore be
checked regularly, and cleaned when necessary. See "Cleaning
and care".
Areas of application
Theatre shoes
can be used as long
43