9.

Printout of Factory Parameters

Open the housing and set jumper „JP1" on main board 172xxx.PCB to access the
service menu.
Connect a standard printer to RS 232 data interface of the moisture analyzer.
Press the ON/OFF key (main switch) to turn on the balance.
Enter main menu
Set index next to parameter
P7 Globals
Enter menu P7 Globals
Enter submenu P7 09 Par. printout
To get a printout of the factory
parameters P0 01 to P0 75 set
Start= 0.01
Stop= 0.75
Confirm your setup with PRINT
Press PRINT to start the printout
16
29.12.04
Setup
P1
C
alibration
P2 GLP
P3 D
ate/Time
P4 Redout
P5 RS-232
P6
Printouts
P7
Globals
MLS...N-SH-e-0812