8.

Off-Center Load Adjustment

Scews to adjust the
off-center load
Zero the display with nothing on the pan. Place a mass of ⅓ of maximum load at the
centre, than at the left, rear, right and front side of the pan, ½ of the way out from the
centre. The readings should agree within ± 5mg.
Using the attached figure. It is necessary to change the screws to make the readings
of the weight equal as it is moved around the pan.
Recheck corner load after 1 hour again.