Filling The Portable Units - CAIRE Companion Series User Manual

Liquid oxygen systems
Table of Contents

Advertisement

Filling the Portable Unit
Fill the Companion Portable unit from the Stationary as follows:
Clean and dry the fill connectors on both the
Stationary and Portable units to prevent freezing and
possible equipment failure.
1. Check the contents indicator on the Stationary to ensure there is
enough liquid oxygen for filling purposes.
For Companion 1000 Transport only: Remove the moisture cup by
rotating the thumb nut (located on the bottom of the moisture cup) a
quarter turn counterclockwise. The moisture pad is held in place with
Velcro. Remove the pad, wring it out, and replace.
2. Using a clean, dry, lint-free cloth, dry the male fill connector (Figure
4) and female fill connector (Figure 8 and Figure 10).
3. Hold the Companion Portable with both hands and position the
contoured case over the matching recessed area in the top of the
Stationary (Figure 11).
Figure 11: Positioning the Portable Unit for Filling
B-701417-00 Rev D
15