"F16" - GRAVITY ZONE OF USE (*)
Through this step one selects the acceleration value of use of the instrument:
Manual entry of the g value: one may manually enter the gravitational acceleration value; one must modify the 6
decimal digits of the gravitational acceleration.
In case one enters a wrong g value: the minimum decimal value is suggested (9,75001); a wrong value is any decimal
number that is not between 9,75001 and 9,84999 (inclusive).
"F17" - SELF-SERVICE MODE
off
disabled
on
enabled
See section 6.9.4 for details.
"F18" - PRINTING FORMAT
Press the M+ key to select one of the possible printing format: Normal, List1, List2.
See section 8 for details.
Confirm with TARE or press ZERO to exit without confirming.
"F19" - PRINTOUT HEADER LINES
By pressing TARE one can configure up to 4 lines that will be printed as header of the printout with the List1 and List2
format (see step F18):
- select the line to configure with M+ key;
- press the TARE key to enter: one is asked to enter the header line through the A
- confirm with the TARE key;
- repeat the operation for all the desired lines or press the ZERO key to exit.
"F20" - PRINTOUT FOOTER LINES
By pressing TARE one can configure up to 2 lines that will be printed as footer of the printout with the List1 and List2
format (see step F18):
- select the line to configure with M+ key;
- press the TARE key to enter: one is asked to enter the header line through the A
- confirm with the TARE key;
- repeat the operation for all the desired lines or press the ZERO key to exit.
"F21" - RESET (*)
For use of the manufacturer.