Conductivity calibration
Measurement parameters
Access to all setting menus can be blocked with an access code !
meas. parameter
setting ?
Cell constant cc
The measured value can be calibrated by changing the cell constant to the actual conductivity value
(arrow keys).
Cell constant calibration (cc)
calibration cc
?
PAR1
Permanent display 1
Permanent display 2
only with control
(w = setpoint)
mr = meas. range
cc = cell constant
Possible values
Initial value
Increment
depending on
0.0001/cm
sensor type
0.001/cm
used
0.01/cm
calibration cc
probe in buffer!
accept value?
1538 µS/cm 8.54
calibration cc
cc: 8.50 /cm
Positive values of setting variable: Increase conductivity
Negative values of setting variable: Reduce conductivity
2000 µS/cm
mr :
8.54/cm
cc :
1106 µS/cm 25,0 °C
Lower value
0.0060/cm
0.150/cm
1.50/cm
calibration cc
cc: 8.54 /cm
cond: 100 µS/cm
D1C2-cond.-011-GB
mr 2 : 2000 µS/cm
cc 2 : 8.54 /cm
1106 µS/cm 25,0 °C
D1C2-cond.-030-GB
Upper value
Remarks
0.1499/cm
cc can be adjusted for
1.499/cm
all mr over the
12.00/cm
complete area
please wait!
D1C2-cond.-040-GB