Complete Operating Menu / Description
Calibration of zero point (zp)
cali. zero point ?
PAR1
To determine the precise zero point (zp) of the sensor, hold the sensor in the air and select the second
menu option. The DULCOMETER
Enter when the value is constant.
The setting menu is inactive when "PAR2" appears.
During the calibration, the metering is reduced to the set basic load and control stops. The limit value
monitor and the fault diagnosis system are reset. The standard signal of the "measured value" output is
frozen.
Cell constant calibration (cc)
calibration cc ?
PAR1
To determine the precise cell constant (cc) of the sensor, place the sensor in a calibration solution
(sample 1) with a known conductivity and select the second menu option. The DULCOMETER
displays the conductivity which it has calculated with the current parameters. Press Enter when the value
is constant.
Select the next menu option and enter the conductivity of the calibration solution (arrow keys). Press Enter.
The D1C displays the recalculated cell constant and saves the value.
The setting menu is inactive when "PAR2" appears.
During the calibration, the metering is reduced to the set basic load and control stops. The limit value
monitor and the fault diagnosis system are reset. The standard signal of the "measured value" output is
frozen.
Initial value
Conductivity
Measured value
Calibration
solution (LF)
cali. zero point
probe in air!
accept value?
12 µS/cm
®
D1C displays the conductivity without zero point correction. Press
calibration cc
probe in buffer!
accept value?
1538 µS/cm 8.54
calibration cc
cc: 8.50 /cm
Possible values
Increment
Lower value
0.1 µS/cm
1 µS/cm
0.01 mS/cm
0.1 mS/cm
1 mS/cm
please wait!
calibration cc
cc: 8.54 /cm
LF: 100 µS/cm
Upper value
0 µS/cm
200 µS/cm
0 µS/cm
2000 µS/cm
0 µS/cm
20 mS/cm
0 µS/cm
200 mS/cm
0 µS/cm
2000 mS/cm
D1C2-cond.-039-GB
please wait!
D1C2-cond.-043-GB
®
D1C
Remarks
Measuring range 200 µS/cm
Measuring range 2000 µS/cm
Measuring range 20 mS/cm
Measuring range 200 mS/cm
Measuring range 2000 µS/cm
15
Need help?
Do you have a question about the DULCOMETER D1C Type D and is the answer not in the manual?