For example, you could set Marker 1 to automatically track the minimum VSWR values in the
trace: «Marker1» «Tracking» «Peak Min» «VSWR»; and Marker 2 to track the crossovers on
the 50-ohm impedance value: «Marker2» «Tracking» «Cross Any» «Z» «50.0».
You could also program the unit to detect the bandwidth by setting «Marker1» «Tracking»
«Cross Down» «VSWR» «2.0»; and «Marker2» «Tracking» «Cross Up» «VSWR» «2.0».
Navigator B will be used to move to the different detection points, except for the Max and Min
tracking modes where logically there is only a single detection point.
The tracking mode for each marker is shown in the markers information window. This
information is displayed in red if either the data is not available or if the tracking condition cannot
be resolved; otherwise it is displayed in green.
The screenshot below shows Marker1 tracking the minimum value of VSWR and Marker 2,
tracking all |Z| crossing at 50-ohms:
The «Info» option displays detailed information of the readings at the marker position, including
the difference in frequency between both markers. One possible use of this feature is
determining the VSWR bandwidth of an antenna.