Advertisement

5.
The User Interface
Start, pause, resume, stop and abort automatic tissue processing cycles
40
START button
To immediately start processing or to insert
delayed starting parameters for automatic pro-
cessing after having selected a program.
PAUSE/CONT button
To pause and resume an automatic processing
cycle.
When the function is activated, the display reads
"PAUSE". In addition, 5 minutes after last pressing
a button, a double sound signal will remind the
user that processing is still paused. This signal
will repeat itself every 5 minutes until the auto-
matic processing cycle is resumed.
When automatic processing is paused, the but-
tons for manual processing become functional,
so that e.g. the tissue basket can be lifted out
of the station to add tissue samples or to unload
them for further/special processing.
The pause period is terminated by pressing the
"PAUSE/CONT" button. The processing cycle is
resumed and the remaining immersion time for
that station is completed without interruption.
STOP button
To immediately stop the carousel up and down
movement (press once) or to abort a processing
cycle in progress (press twice).
The display reads "STOP?".
Instructions for Use V 3.0 RevG, English – 07/2016