4.2
Installation site requirements
Because of the quantity of solvents
used for processing specimens and the
solvent concentrations that build up,
we recommend installing instrument
versions without a fume control system
or active carbon filter in the labora-
tory's fume hood.
780 mm
Leica TP1020
820 mm
Fig. 12
The instrument may not be operated in
hazardous locations.
The installation site has to fulfill the following
conditions:
• Stable and plane installation surface.
• Minimum installation surface dimensions:
850 x 850 mm,
• Room temperature consistently between +5 °C
and +40 °C,
• Relative humidity should not exceed 80%.
Necessary conditions for electrical connection
- Grounded wall outlet near enough so it can
be reached with the power cable supplied
together with the instrument.
If you wish to install the basket holder
for a second tissue basket, you should
do that now.
For mounting instructions, see
ter 13 "Optional
4.
Instrument Setup
Chap-
Accessories".
29