12.2 Wax bath

12.2.1 Altering the standard working temperature

Factory-set standard working temperature is 65 °C. Wax bath heating will automatically switch
on when the actual temperature of the paraffin is 5 °C below set working temperature.
When working with paraffin that has a melting point below 58 °C, the instrument working
temperature can be readjusted with the corresponding setting screw.
Leica TP1020
• Use a screw driver to turn the setting screw
(1) to the desired value.
If you find that the paraffin does not melt com-
pletely after lowering the working temperature,
slightly readjust again.
1
Fig. 24
12.
Maintenance
79