4 - Pre-Installation Planning

Pre-Installation Planning
Before your booth is installed, you should observe the following considerations.
Shock hazard.
Disconnect power before
servicing.
WARNING
Always wear protective eyewear when unit
is on.
Failure to may result in severe burns or
longterm injury to the eyes.
CAUTION
Use of a voltage source above 230V AC
(with unit running) may prevent proper oper-
ation of the unit and could cause damage
and void the warranty.
CAUTION
Air from the room is used to cool the unit.
Maximum ambient room temperature
should be 80°F. Place your booth no closer
than 6" from walls. Make sure nothing
obstructs the airflow into or out of the fan
openings. A poorly ventilated room may
cause the unit to become hot and cause
discomfort to the user.
DANGER
• The SolarForce® 652V requires a dedicated
circuit capable of providing 70A, 1-phase
service (2 wire plus ground) at 220VAC. This
unit must be hardwired directly to a junction
box on a dedicated circuit. We recommend
connection by a professional electrician.
Consult local and/or national codes for proper
circuit breaker and wire gauge.
• IMPORTANT! Operating voltage must be
below 230V AC or may require a Buck
Booster.