Questionable Status Group

The Questionable Status registers record signals that indicate abnormal operation of the dc source. As
shown in figure 7-7, the group consists of the same type of registers as the Status Operation group. The
outputs of the Questionable Status group are logically-ORed into the QUEStionable summary bit (3) of
the Status Byte register.
Register
Command
STAT:QUES:COND?
Condition
STAT:QUES:PTR <n>
PTR Filter
STAT:QUES:NTR <n>
NTR Filter
STAT:QUES:EVEN?
Event
STAT:QUES:ENAB <n>
Enable

Standard Event Status Group

This group consists of an Event register and an Enable register that are programmed by Common
commands. The Standard Event event register latches events relating to instrument communication status
(see figure 7-7). It is a read-only register that is cleared when read. The Standard Event enable register
functions similarly to the enable registers of the Operation and Questionable status groups.
Command
Action
*ESE
programs specific bits in the Standard Event enable register.
*PSC ON
clears the Standard Event enable register at power-on.
*ESR?
reads and clears the Standard Event event register.
The PON (Power On) Bit
The PON bit in the Standard Event event register is set whenever the dc source is turned on. The most
common use for PON is to generate an SRQ at power-on following an unexpected loss of power. To do
this, bit 7 of the Standard Event enable register must be set so that a power-on event registers in the ESB
(Standard Event Summary Bit), bit 5 of the Service Request Enable register must be set to permit an SRQ
to be generated, and *PSC OFF must be sent. The commands to accomplish these conditions are:
*PSC OFF
*ESE 128
*SRE 32

Status Byte Register

This register summarizes the information from all other status groups as defined in the IEEE 488.2
Standard Digital Interface for Programmable Instrumentation. See Table 7-1 for the bit configuration.
Command
Action
*STB?
reads the data in the register but does not clear it (returns MSS in bit 6)
serial poll
clears RQS inside the register and returns it in bit position 6 of the response.
A register that holds real-time status of the circuits being monitored. It is a
read-only register.
A positive transition filter that functions as described under
STAT:QUES:NTR|PTR commands in chapter 8. It is a read/write register.
A negative transition filter that functions as described under
STAT:QUES:NTR|PTR commands in chapter 8. It is a read/write
register.
A register that latches any condition that is passed through the PTR or NTR
filters. It is a read-only register that is cleared when read.
A register that functions as a mask for enabling specific bits from the Event
register. It is a read/write register..