Performing The Calibration Procedure - Agilent Technologies 66319B User Manual

Mobile communications dc source
Hide thumbs Also See for 66319B:
Table of Contents

Advertisement

B - Performance, Calibration, and Configuration

Performing the Calibration Procedure

NOTE:
The calibration procedure can only be performed from the front panel or using the SCPI
language commands.
Table B-1 lists the equipment required for calibration. Figure B-1 shows the test setup.
You do not have to do a complete calibration each time. If appropriate, you may calibrate only the
voltage or current and proceed to "Saving the Calibration Constants". However, the voltage or current
calibration sequence must be performed in its entirety. The following parameters may be calibrated:
voltage programming and measurement
current programming and measurement
1 A and 0.02A range measurements
ac current measurement
resistance programming
internal DVM
Front Panel Calibration Menu
The Entry keypad is used for calibration functions.
Shift
Cal
Press this key to access the calibration menu.
Display
CAL ON <value>
CAL OFF
CAL:LEV <char>
CAL:DATA <value>
CAL:VOLT
CAL:VOLT2
CAL:RES
CAL:VOLT:PROT
CAL:CURR
CAL:CURR2
CAL:CURR:MEAS:LOW
CAL:CURR:MEAS:R3
CAL:CURR:MEAS:AC
CAL:DVM
CAL:SAVE
CAL:PASS <value>
DATE <char>
Notes:
value = a numeric value
char = a character string parameter
Use
Use
Use
154
Command Function
Turns calibration mode on when the correct password
value is entered.
Turns calibration mode off
Advance to next step in sequence (P1 or P2).
Enter an external calibration measurement.
Begin voltage calibration sequence
Begin output 2 voltage calibration sequence
Begin resistance calibration
Not required
Begin high range current calibration sequence
Begin output 2 current calibration sequence
Begin 0.02A range current measurement calibration
Begin 1A range current measurement calibration
Begin ac current calibration sequence
Begin DVM calibration sequence
Saves the calibration constants in non-volatile memory.
Set new calibration password.
Displays the calibration date. (0 if none is supplied)
and
to scroll through the menu commands.
and
to scroll through the menu parameters.
and
to select a digit in a numeric entry field.