Front Panel Calibration Procedure - Agilent Technologies 66319B User Manual

Mobile communications dc source
Hide thumbs Also See for 66319B:
Table of Contents

Advertisement

Front Panel Calibration Procedure

These procedures assume you understand how to operate front panel keys (see chapter 5). Make sure the
sense terminals are directly jumpered to the output terminals.
Enable Calibration Mode
Action
1.
Reset the unit by selecting Output, scrolling to *RST and pressing Enter.
2.
Press Output On/Off to enable the output.
3.
To begin calibration press Shift Cal, scroll to CAL ON and press Enter.
4.
Enter the calibration password from Entry keypad and press Enter. lf the
password is correct the Cal annunciator will come on.
If CAL DENIED appears, then an internal switch has been set to prevent the
calibration from being changed.
lf the password is incorrect, an error occurs. If the active password is lost, the
calibration function can be recovered by setting an internal switch that defeats
password protection (contact the factory for details).
Voltage Programming and Measurement Calibration
Action
5.
Connect the external DMM (in voltage mode) directly to output 1.
Do not connect the load resistor or current shunt. Select output 1.
6.
Press Shift Cal, scroll to CAL VOLT, and press Enter.
7.
Press Shift Cal, scroll to CAL LEV, and press Enter to select the first
calibration point.
Press Shift Cal, scroll to CAL DATA, press Enter Number, and enter the
8.
voltage value displayed on the DMM. Press Enter.
9.
Press Shift Cal, scroll to CAL LEV, use
calibration point), and press Enter.
10.
Press Shift Cal, scroll to CAL DATA, press Enter Number, and enter the
second voltage value displayed on the DMM. Press Enter.
Agilent 66319B/D Output 2 Voltage Programming and Measurement Calibration
Action
11.
Connect the external DMM (in voltage mode) directly to output 2.
Do not connect the load resistor or current shunt. Select output 2.
12.
Press Shift Cal, scroll to CAL VOLT2, and press Enter.
13.
Press Shift Cal, scroll to CAL LEV, and press Enter to select the first
calibration point.
Press Shift Cal, scroll to CAL DATA, press Enter Number, and enter the
14.
voltage value displayed on the DMM. Press Enter.
15.
Press Shift Cal, scroll to CAL LEV, use
calibration point), and press Enter.
16.
Press Shift Cal, scroll to CAL DATA, press Enter Number, and enter the
second voltage value displayed on the DMM. Press Enter.
Performance, Calibration, and Configuration - B
to scroll to P2 (the second
to scroll to P2 (the second
Display
*RST
00.003V
0.0006A
CAL ON 0.0
CAL DENIED
OUT OF RANGE
Display
CAL:VOLT
CAL:LEV P1
CAL:DATA 0.00
CAL:LEV P2
CAL:DATA 0.00
Display
CAL:VOLT2
CAL:LEV P1
CAL:DATA 0.00
CAL:LEV P2
CAL:DATA 0.00
155

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

66319d66321b66321d

Table of Contents