Manual Media Calibration

MP Nova Mark II User Guide
If the calibration is not successful, repeat the procedure. If the Quick Media
calibration continues to fail, proceed to
page
43.
The Manual Media Calibration is intended for use when Quick Media
Calibration printer has failed to detect the start of each label. This procedure
performs a complete recalibration of the sensors and will optimize the printer to
your media. In some instances you may need to perform a Quick Media
Calibration after the Manual Media Calibration to further optimize the printer's
sensor.
Sensor calibration is needed to set either the black mark or the gap value of the
media sensor on the printer. The Manual Media Calibration can be performed
using the NETira CT configuration utility, proceed to
Utility (NETira CT)" on page
1 Launch the NETira CT configuration utility and query (connect) to the printer.
2 Click the Manual Media button to start the manual calibration process.
Follow the on-screen instructions.
3 Select the Interlabel/Gap radio button, then click OK.
4 Load the Stock by placing the face of the media over the sensor. Lower the
cover and click OK.
Chapter 5 — Printer Operation
"Manual Media Calibration" on
"Printer Configuration
31.