A1.2.8 Spo - EDAN M3 Service Manual

Vital signs monitor
Hide thumbs Also See for M3:
Table of Contents

Advertisement

Adult Mode
Pediatric Mode
Neonatal Mode
Cuff Pressure measuring Range
Pressure Resolution
Pressure Accuracy
Mean error
Maximum Standard deviation
Entire Measuring Period
Overvoltage protection

A1.2.8 SpO

2
Measuring Range
Alarm Range
Resolution
Accuracy
Adult (including Pediatric)
Neonate
Pulse Rate
Measuring and Alarm Range
Resolution
Accuracy
Anti-low-perfusion
M3 Vital Signs Monitor Service Manual
SYS
DIA
MAP
SYS
DIA
MAP
SYS
DIA
MAP
0~280mmHg
1mmHg
±3 mmHg
±5 mmHg
≤8 mmHg
20~45s typical (depend on HR/motion disturbance)
Dual Overvoltage protection
Adult
Pediatric
Neonatal
0 ~ 100 %
0 ~ 100 %
1 %
±2 digits(70%~100% SpO
Undefined(0~70% SpO
±3 digits(70%~100% SpO
Undefined(0~70% SpO
15 ~ 254 bpm
1 bpm
± 3bpm
Under Motion Condition, ±5 bpm
Strong Anti-low-perfusion, 0.075 ~ 0.1%
- 48 -
30~270mmHg
10~220mmHg
20~235mmHg
30~235mmHg
10~220mmHg
20~225mmHg
30~135mmHg
10~110mmHg
20~125mmHg
297±3mmHg
240±3mmHg
145±3mmHg
2
2
2
2