To replace the rubber bag in the cuff, first place the bag on top of the cuff so that the rubber tubes
line up with the large opening on the long side of the cuff. Now roll the bag lengthwise and insert
it into the opening on the long side of the cuff. Hold the tubes and the cuff and shake the
complete cuff until the bag is in position. Thread the rubber tubes from inside the cuff, and out
through the small hole under the internal flap.
Disposable Blood Pressure Cuffs
Disposable cuffs are intended for one-patient use only. Do not use the same cuff on any other
patient. Do not sterilize or use autoclave on disposable cuffs. Disposable cuffs can be cleaned
using soap solution to prevent infection.
NOTE:
For protecting environment, the disposable blood pressure cuffs must be recycled or
disposed of properly.
7.1.2 SpO
Sensor Maintenance and Cleaning
2
WARNING
Before cleaning the monitor or the sensor, make sure that the equipment is switched off
and disconnected from the power line.
WARNING
Do not subject the sensor to autoclaving.
Do not immerse the sensor into any liquid.
Do not use any sensor or cable that may be damaged or deteriorated.
For cleaning:
n Use a cotton ball or a soft mull moistened with hospital-grade ethanol to wipe the surface of
M3 Vital Signs Monitor Service Manual
Figure7-1 Replace Rubber Bag in Cuff