2.0 Performance Parameters and Features

Both the interior and exterior of the DH are constructed of 18 to 16-gauge steel.
The interior is highly polished type 304 L 16 gage stainless steel, using crevice-free construction.
All exposed edges are deburred to insure no sharp edges.
The exterior is cold rolled 16 & 18 gauge steel finished in a powder coated polyurethane finish, which is resistant to
chemicals and easily cleaned using mild household detergents.
The DH's microcomputer temperature control system has two temperature sensors located inside the chamber at
the back wall.
The temperature sensors compare the values to a set point and execute a differential control algorithm that
energizes a solid-state switch, supplying power to the heaters.
Easily removable inner shelves and rails for sterilization.
Up to 21 shelves can be placed inside the chamber (4 shelves are shipped with each DH).
Space-age high temperature material is used to insulate the inner chamber walls.
Foil heaters are directly attached to the chamber walls and the top and bottom surfaces of the inner chamber.
Most electronics, motors, pumps, and valves are fully accessible from the back of the unit.
A CO
sample port is provided on the front panel just below the door to check the concentration of CO
2
chamber.
The CO
percentage is controlled by a solid-state gas infrared sensor using a single light source with dual wave length
2
detection, providing accurate monitoring of CO
Automatic recovery of the CO
The outer door includes an internal radiant heater in order to minimize condensation on the inner glass door.
A magnetic outer door gasket helps to insure a tight seal against the cabinet.
The inner glass door is 3/16" tempered with smooth-ground edges and seals are tight against a silicone rubber
gasket.
The door latch is cam action.
A momentary switch monitors the door position, the position being opened or closed.
All control electronics are protected with a circuit breaker that may trip at 110% of loading rating but will trip at
145% of load rating in less than 2 seconds.
Should the circuit breaker open (pop-out button will appear), merely depress to reset.
The Incubator has factory-installed adjustable leveling legs to compensate for uneven laboratory surfaces.
OM0211
Rev 3 March 2013
.
2
level, after a 1minute door opening, to 5.0 + 0.5% is within a 5 minute period.