7.4 Electrical

The electrical supply circuit to the Incubator must conform to all national and local electrical codes.
Consult the serial-data plate, located at the front of the right side of the Incubator, for voltage, cycle, phase, and ampere
requirements before making connection. Plug the power cord securely into a grounded power source. VOLTAGE SHOULD
NOT VARY MORE THAN 5% FROM SERIAL PLATE RATINGS. Have a qualified technician check with the power source with a
properly rated volt meter if needed. A separate branch circuit is recommended to prevent possible loss of product due to
overloading or failure of other equipment on the same circuit. A SURGE PROTECTOR IS STRONGLY RECOMMENDED to
avoid power-related faults.

7.5 Air Inlet Connection

The air inlet is installed at the Factory and consists of flexible hose attached to bulkhead fittings in the bottom of the front
panel and can be found under the hinged Valance that is beneath the door. One 50mm polypropylene .3-micron HEPA filter
is connected to the hoses. The filter is labeled "Air filter". To open the valence the outer door must be open. Reach in
under the front of the valence and pop it open by pulling it toward yourself.
See BCD-15180
CAUTION:

7.6 CO2 Connection

CAUTION
!
This Incubator is designed to be operated with CO
condition. Gases other than CO
cylinder and are equipped with a CGA 320 outlet valve. Check the gas cylinder for the proper identification labels.
Do not use CO
gas cylinders equipped with siphon tubes. A siphon tube is used to extract liquid CO
2
which can damage the pressure regulator. Consult with your gas supplier to ensure that the CO
siphon tube.
OM0211
Rev 3 March 2013
This is a free (open air atmospheric) air supply. DO NOT connect to a pressurized air source.
High concentrations of CO
Install Incubator in a well-ventilated area.
should not be connected to this equipment. CO
2
gas can cause asphyxiation!
2
gas only. Connecting a flammable or toxic gas can result in a hazardous
2
Page 15 of 48
gas cylinders have a UN1013 label on the
2
from the cylinder
2
cylinder does not contain a
2