Automatic Filter Change Notification; Responding To The Notification - NuAire NU-5800 Operation And Maintenance Manual

Air-jacketed dh invitrocell co2 incubator
Hide thumbs Also See for NU-5800:
Table of Contents

Advertisement

10.3 Automatic Filter Change Notification

The Capsule filter located in the Sensor bay (see BCD- 15634 for location)
should be checked average every 2 years to see if it should be changed. The
Filter material is normally a white material. Discoloration of the filter material
is an indicator that it should be changed. Retro Fit Kit RF566 should be ordered
for the replacement filter. A timed notice triggers a Maintenance Notification
button in the Main Screen when this timer counts down to zero. Pressing the
Maintenance Required button produces a screen indicating that the timer for
filter life has expired.

Responding to the notification:

From the main screen press the NuAire Logo / Service settings / General
Option / Settings / Filter Maintenance to access the Filter Maintenance
Adjustments screen. In this screen you can extend the alarm by 1 month
by pushing the Extend? button. Reset the notification timer from 1 month
to 48 months by sliding the green button right or left on the scale. The
number of months showing will change as the button is moved. Setting
the timer to 0 months deactivates the notification. After extending or
resetting the timer you can return to the main screen by following the
screen prompts.
OM0211
Rev 3 March 2013
Page 38 of 48