Maintaining Your Dh; Chemical Decontamination Of The Incubator Chamber; Shutting Down The Incubator; Filters Maintenance - NuAire NU-5800 Operation And Maintenance Manual

Air-jacketed dh invitrocell co2 incubator
Hide thumbs Also See for NU-5800:
Table of Contents

Advertisement

10.0

Maintaining Your DH

DH Chamber
The chamber maintenance is up to the discretion of the owner and the extent of cleanliness and sterility desired.
The shelves and bracket supports are all removable and autoclavable.
The interior should be wiped down with an appropriate disinfectant such as 70% ISOPROPYL ALCOHOL or equivalent.
Note: DO NOT USE ANY CHLORINATED OR HALOGEN MATERIALS IN THE CHAMBER.
SUCH MATERIAL IS HARMFUL TO THE POLISHED STAINLESS STEEL.
The humidity pan should also be sterilized and the water changed regularly to assure sterility.
A small amount of copper sulfate may be added to the humidity pan to inhibit bacterial growth.
CO
Supply Filter
2
P/N X-980385 (50 mm Disk, Uni-directional In-Line, Dry)*
The CO
Supply Filter should be replaced every fifth empty CO
2
Sensor Bay Capsule Filter
P/N RF566 (Capsule, Uni-Directional In-Line, Wet)
The CO2 Sensor Filter should be replaced EVERY TWO YEARS to assure optimum performance.
A visual check should be performed during CO2 sensor calibration to assure filter integrity.
Remove sensor housing in back of unit cover to perform visual check.
See section 10.3 for automatic notification for checking this filter.
Outlet port is on flat top side.
Air Supply Filter
P/N X-980385 (50 mm Disk, Uni-directional In-Line, Dry)*
The Air Supply Filter should be replaced every fifth empty CO
Air Pump Filters
P/N X-980366 (50mm Disk, Uni-Directional In-Line, Wet)*
The Air Pump Filters should be replaced EVERY TWO YEARS to assure optimum performance.
A visual check should be performed during CO
Remove the sensor housing cover to perform visual check.
* Note: The word "In" on the outer rink of the body indicates the inlet side of the filter
and should be installed toward the gas supply.

10.1 Chemical Decontamination of the Incubator Chamber

To chemically decontaminate NuAire Incubator inner chambers, users may use the traditional formaldehyde or Chlorine
Dioxide in gas form, or Vapor Phased Hydrogen Peroxide.
All three of the chemicals are compatible to all parts within NuAire Incubator chambers.
The incubator should be turned off during these procedures
Note: As stated previously, the chamber and components can also be wiped down with a
70% solution of Isopropyl Alcohol for cleaning and decontamination.

10.2 Shutting down the Incubator

Prior to shutting down the Incubator open the inner and outer doors and remove the water pan.
Leave doors open for at least 5 minutes prior to shutting it off.
This will purge the chamber, circulating system and the sensors of humidity that could condense and cause faulty
readings when the Incubator is turned back on.
Be sure to empty the water pan prior to putting it back into the chamber if the Incubator is going to be shut off for any
length of time.
OM0211
Rev 3 March 2013
tank or when the filter is visibly discolored (yellow-brown).
2
tank or when the filter is visibly discolored (yellow-brown).
2
sensor calibration to assure filter integrity.
2