Error Indicator Troubleshooting Guide - NuAire NU-5800 Operation And Maintenance Manual

Air-jacketed dh invitrocell co2 incubator
Hide thumbs Also See for NU-5800:
Table of Contents

Advertisement

Displayed Error Code
Temp Over Set Point,
Temp Under Set Point,
Failed Sensor Differential
Sensor Bay Temp Low
CO2 Over Set Point
CO2 Under Set Point
Cal Inject Failed
Door Ajar/rapid beeping
"Power Interuption" Text
replacing NuAire Logo
Corrupted Memory
Data Write Failure
12 VDC Power Failure
5 VDC Power Failure
Zero Cross Failure
OM0211
Rev 3 March 2013

Error Indicator Troubleshooting Guide

Code Description
Temperature System
-Temperature over setpoint
-Temperature time out error
during normal running and in
-Sensor temperature (differential)
error normal running.
-Occurs when difference between
sensors exceeds 4°C.
CO2 Sensor Shroud Heater
-Low shroud temperature
- CO
over setpoint
2
- CO
time out error
2
- Cal inject calibration failed.
Not enough increase in the CO
reading after gas was injected.
General Indicators
- Inner glass door is not closed or the
door switch needs attention.
- Interruption of power
Corrupted memory was detected
during boot up of the incubator
A change in parameters failed to write
to memory
Low voltage detected on the 12 VDC
power circuit
Low voltage detected on the 5 VDC
power circuit
The Power Triacs control circuit not
working
Check temperature sensor calibration.
Faulty TRIAC, replace control board.
Check temperature sensor calibration
Replace fuse.
Faulty TRIAC, replace control board
Faulty chamber heater contact NuAire Technical Service.
Door/Perimeter heater needs to be increased with a high
temperature set-point in a low ambient temperature
Check temperature sensor calibration
Check connection on control board
One or both temp sensors faulty, replace
Check control board output to relay, replace board if faulty.
Check heater relay for function, replace if faulty
Check both shroud heaters for function, replace if faulty.
CO
System
2
Perform CO
sensor calibration
2
Check injection solenoid for leaking valve
Check sensor and disk filter for condensation
Check CO
gas supply - inline gas filters, CO
2
pressure, CO
Run Cal Inj. Calibration (see Section 9.3.3)
Check/replace CO
Check for leaks in chamber
- inner door gasket, chamber blower fan shaft seal
Check for leaks in air pump and hosing
Check gas supply then run calibration again
Call NuAire Technical Services if error persists.
2
Close and latch inner glass door.
Confirm placement on the Control Board
Check door switch, if faulty, replace.
Press any key to dismiss the power interruption notice.
Replace Control Board
Replace Control board
Check output @ Power Supply-If good could be Control Bd
Check power supply connection to Control board
DC power supply failed - replace
Check output @ Power Supply-If good could be Control Bd
Check power supply connection to Control board
DC power supply failed - replace
Check for loose line voltage connection to control board
Replace fuse with slow blow fuse of the correct amp rating
as specified on the wiring diagram
Replace Control Board
Page 40 of 48
Checks and Corrections
sensor function
2
gas supply tanks
2
gas tank
2