Chamber Temperature Calibration - NuAire NU-5800 Operation And Maintenance Manual

Air-jacketed dh invitrocell co2 incubator
Hide thumbs Also See for NU-5800:
Table of Contents

Advertisement

9.0 Calibration
Proper calibration of the incubator involves four parameters: chamber temperature, door temperature, perimeter temperature,
and CO
sensor. The first three, chamber, door, and perimeter temperature should be completed and stabilized before any CO
2
sensor calibration is performed. Below, each calibration procedure is described in detail. For the best results, follow the
procedure carefully, and if the desired result is not achieved, try procedure again from the start.

9.1 Chamber Temperature Calibration

The DH's TEMPERATURE CALIBRATION MUST BE PERFORMED WITHIN 1°C OF THE PLANNED OPERATING TEMPERATURE.
Normally, 37.0°C is the most common set point. To initiate the procedure, turn on the incubator via the power switch on
the back panel. The default set point is 37.0°C; enter "Environmental Set Points" and adjust if desired Let the unit stabilize
for 8 to 12 hours. Use an independent calibrated instrument to check and calibrate the temperature.
Below is a description of calibration with a glass thermometer as one example for temperature calibration.
At the beginning of this procedure, set a calibrated glass thermometer in a glass beaker filled with water resting on a shelf
in the middle of the chamber. Do not place the glass beaker on the bottom of the chamber because it will result in a slightly
higher temperature reading due to the heater located on the chamber bottom. Placing the thermometer in glass beaker on
the middle shelf will give the most accurate results for calibration. The chamber should be humidified to avoid false low
readings due to evaporation of water from the flask. An accurate digital thermometer with a type K thermal couple could
also be used.
When the unit has stabilized at the operating temperature, perform the following calibration procedure.
Allow Incubator to stabilize at its given temperature set point in run mode.
Enter "Service Settings" by means of your provided password.
Press the Temperature Calibration button.
Determine actual temperature within chamber by reading temperature measurement instrument.
Select the number digits below the "Current Calibrated Sensor Reading" blue text button
Input the current value displayed on your temperature measurement instrument and exit calibration.
Note: THE DISPLAY DOES NOT UPDATE IMMEDIATELY; IT TAKES 2-3 SECONDS
The chamber temperature calibration is complete. Let unit stabilize for 8 to 12 hours. If the chamber temperature (actual
thermometer) still does not match the display, perform the above procedure again. In some cases it might be necessary to
calibrate several times to achieve a stable condition due to ambient conditions of temperature and humidity within the
laboratory.
OM0211
Rev 3 March 2013
Page 33 of 48
2

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Nu-5800e

Table of Contents