Installation
October 2016
1. For adapters with BSP/G threads, place a gasket on top of the tank flange.
2. For adapters with NPT threads, use anti-seize paste or PTFE tape according to your site procedures.
3. Lower the transmitter and probe into the tank.
4. Loosen the nut that connects the transmitter head to the probe slightly.
5. Screw the adapter into the process connection.
6. Rotate the transmitter head so the device display faces the desired direction.
7. Tighten the nut. Max torque is 30 Lbft (40 Nm).
8. Continue with the grounding step.
3.6.2

Tank connection with flange

Figure 3-17. Tank Connection with Flange
A. Transmitter head
B.
Nut
C. Bolts
D. Flange
34
A
B
C
D
E
F
G
E.
Probe
F.
Gasket
G. Tank flange