Emerson Rosemount 3308A Reference Manual page 175

Wireless guided wave radar, rosemount 3308 series
Hide thumbs Also See for Rosemount 3308A:
Table of Contents

Advertisement

Configuration Parameters
October 2016
Device Display
Display Mode
Display Variables
Units
The units for length, volume, and temperature are selectable. After appropriate units have been
selected, all configuration parameters and transmitter variables will be expressed in these units.
Length Unit
Volume Unit
Temperature Unit
HART - Variable Mapping
Primary Variable
Secondary Variable
Third Variable
Fourth Variable
HART - Percent of Range
Upper Range Value
Lower Range Value
Upper Sensor Limit
Lower Sensor Limit
Configuration Parameters
The display can be configured to different display modes: Disabled, On
Demand, or Periodic.
Display mode Description
Disabled
The display is always turned off.
On Demand
The display is by default turned off. Selected variable
screens will only appear in the end of the diagnostic
button screen sequence, refer to
screen sequence" on page
Periodic
The display shows selected variable screens in a
periodic sequence. A new screen will appear on each
wireless update.
As default, the level variable will be displayed. If more than one variable is
configured, the display will toggle between the values of the chosen
variables.
Used unit for Level and Interface Level values.
Used unit for Volume values.
Used unit for Electronics Temperature value.
Primary dynamic variable in the HART protocol which will be assigned as a
variable from the device.
Second dynamic variable in the HART protocol which will be assigned as a
variable from the device.
Third (Tertiary) dynamic variable in the HART protocol which will be
assigned as a variable from the device.
Fourth (Quaternary) dynamic variable in the HART protocol which will be
assigned as a variable from the device.
Value for Primary Variable (PV) corresponding to 100% range.
Value for Primary Variable (PV) corresponding to 0% range.
The upper boundary for the range over which the sensor works properly.
The lower boundary for the range over which the sensor works properly.
Reference Manual
00809-0100-4308, Rev BB
"Diagnostic button
72.
163