Machine Connections - Miele PW 6207D Installations Plan

Advertisement

Machine connections:

Electrical
1. Standard voltage (standard version)
connection
Variations in the following countries:
1. Standard voltage (standard version)
Alternative voltage (convertible)
The electrical connection may only be made to an electrical
system provided in accordance with all appropriate local and
national legislation and regulations. In addition, all regulations
of the appropriate utility supply companies and standards
relating to safety, and all applicable valid regulations and
technical standards must be complied with!
Electrical connection is the responsibility of a qualified
electrician.
The machine can be either hard-wired or connected via a
socket in compliance with IEC 60309-1. Wall socket or mains
isolator must be accessible after installation.
It is always recommended to make electrical connection via a
plug and socket so that electrical safety checks, e.g. during
repair or service work, can be carried out easily.
If it is necessary to install a residual current device (RCD) in
accordance with the local regulations, a residual current
device type B must be used.
Equipotential
Connection with male thread
bonding
for washers and nuts
Fasteners such as washers and nuts are not provided.
If necessary, equipotential bonding with good galvanic
contact must be guaranteed in compliance with all applicable
local and national installation codes.
Steam
Steam pressure (Low pressure steam)
connection
Direct
Boiling point
Mass flow (max. delivery)
Heater rating
Steam pressure (high-pressure steam)
Boiling point
Mass flow (max. delivery)
Heater rating
On-site threaded union
Steam valve, filter, non-return valve and steam shut-off valve
must be provided on site.
The steam valve should be installed in such a way as to
prevent any wash liquor from the machine from flowing back
into the valve.
Technisches Datenblatt / Technical datasheet: PW 6207 D
Datum / Date: 24.04.2013
Frequency
Connected load
Fuse rating
Connection cable, min. gauge
Connection cable not supplied (to be ordered
separately)
Frequency
Connected load
Fuse rating
Connection cable, min. gauge
Connection cable not supplied (to be ordered
separately)
Frequency
Connected load
Fuse rating
Connection cable, min. gauge
Connection cable not supplied (to be ordered
separately)
V
3N AC 380 - 415
Hz
50 - 60
kW
4.3
A
3 × 16
mm²
5 × 1.5
V
3 AC 220 - 240
Hz
60
kW
4.3
A
3 × 15
mm²
4 × 1.5
V
3 AC 208
Hz
60
kW
4.3
A
3 × 15
mm²
4 × 1.5
mm
10 × 35
M
10
kPa
50
120
°C
kg/h
22
kW
13
kPa
50 - 400
°C
152
kg/h
52
kW
30
Inch
½" female thread
Seite / Page:
8

Advertisement

loading

This manual is also suitable for:

Pw 6137 d