Description Of The Machine - Miele PG 8528 Installation And Operating Instructions Manual

Washer-disinfector
Table of Contents

Advertisement

Miele washer-disinfectors are designed
to clean, rinse, disinfect
(thermal/thermochemical) and dry
re-usable medical devices, laboratory
equipment and instrumentation.
Whatever items are being processed,
always comply with the particular
manufacturer's instructions (EN ISO
17664).
Typical applications include:
– surgical instruments,
– endoscopic instruments,
– intensive care and anesthesiology
equipment,
– baby bottles and nipples,
– supply and disposal containers,
– OR shoes,
or
– laboratory equipment for research
and production,
– laboratory equipment for analytical
and preparation applications,
– microbiology and biotechnology
labware.
Labware can include anything from
evaporation dishes to centrifuge tubes.

Description of the machine

In this operating manual, the general
term labware is used to cover a range
of equipment and items.
For standardized results, it is preferable
to machine-wash lab instruments and
equipment.
When disinfection is required to protect
staff and/or patients, the preference is
for thermal disinfection, e. g., with the
DESIN vario TD process.
According to the A
in EN ISO 15883-1, thermal disinfection
occurs at 80°C (+ 5 °C, - 0 °C) with
10 min holding time (A
(+ 5 °C, - 0 °C) with 5 min holding time
(A
3000), depending on the
0
disinfection result required. The
efficacy standard A
inactivation of HBV.
Regional and/or other official directives
may apply (e. g., §18 of the IfSG for
Germany).
The cleaning parameters should always
be optimally matched to the type of
soiling and labware involved.
The processing chemicals used also
should be matched to the particular
cleaning needs / analysis / analytical
methods involved.
concept described
0
600), or at 90°C
0
3000 includes the
0
9