10. Watch the messages on the screen. When the devices are paired, messages similar to the following
appear on the screen ending with "Pairing Succeeded":
If pairing
Then:
is:
Successful 1. Press the back icon. The Blood Pressure screen appears with the serial number of the
paired blood pressure monitor listed:
2. Remove the blood pressure cuff.
Not
1. Press the back icon. The Blood Pressure screen appears.
Successful
2. Go to step 4 of Pairing a Blood Pressure Unit and try again.
3. If you are using the A&D Blood Pressure Unit, UA-767 PBT-Ci, and pairing is still
not successful after a second attempt, then hold the start button down on your A&D
unit until you see "uPr" displayed on the screen.
4. Briefly press the start button again while "uPr" is still displayed.
5. When the monitor displays "End" you have reset your Bluetooth pairing settings.
6. Please go to step 4 of Pairing a Blood Pressure Unit and try again.